| 04416A1 | SERVOPRO NanoChrome O2 Background | 
| includes report software, packaging | |
| Impurities options | |
| N2 Range 0-250ppb LDL 0.3ppb | INJ/HC with 2 valves | 
| Ar Range 0-250ppb LDL 0.5ppb | INJ/HC with 2 valves | 
| H2 Range 0-250ppb LDL 0.8ppb | INJ/HC with 2 valves | 
| CO Range 0-250ppb LDL 0.5ppb | BF/HC with 2 valves | 
| CO2 Range 0-250ppb LDL 0.8ppb | BF/HC with 2 valves | 
| CH4 Range 0-250ppb LDL 0.8ppb | BF/HC with 2 valves | 
| CO2/CH4 Range 0-250ppb LDL 0.8ppb | BF/HC with 2 valves | 
| NMHC Range 0-250ppb LDL 0.8ppb | BF with 1 valve | 
| Chassis options | |
| Master Chassis | Max. 2 channels with 4 valves | 
| PC+ Slave chassis | 2 BF/HC channels | 
| Master +Slave1 Chassis | Max. 4 channels with 8 valves | 
| Master +Slave 1 +Slave 2 Chassis | Max. 6 channels with 12 valves | 
| Aux.oven | O2 trap only for N2 measurement | 
| Carrier Gas | Helium | 
| Power | 115 VAC, 50-60 Hz | 
| 220 VAC, 50-60 Hz | |
| Gas Connectors (to match accessories fittings) | 1/8" VCR | 
| Serial port | RS-232 | 
| Remote Support | VNC | 
| Net Support | |
| Internal Dilution System | Not Required( > 1ppm) | 
| Required( One per Chassis) | |
| 04416A1 | SERVOPRO NanoChrome H2 Background | 
| includes report software, packaging | |
| Impurities options | |
| N2 Range 0-250ppb LDL 0.3ppb | INJ/HC with 2 valves | 
| Ar Range 0-250ppb LDL 0.5ppb | INJ/HC with 2 valves | 
| CO Range 0-250ppb LDL 0.5ppb | BF/HC with 2 valves | 
| CO2 Range 0-250ppb LDL 0.5ppb | BF/HC with 2 valves | 
| CH4 Range 0-250ppb LDL 0.5ppb | BF/HC with 2 valves | 
| CO2/CH4 Range 0-250ppb LDL 0.5ppb | BF/HC with 2 valves | 
| CO2/CH4/NMHC Range 0-250ppb LDL 0.5ppb | BF/HC/SEL with 3 valves | 
| Chassis options | |
| Master Chassis | Max. 2 channels with 4 valves | 
| PC+ Slave chassis | 2 BF/HC channels | 
| Master +Slave1 Chassis | Max. 4 channels with 8 valves | 
| Master +Slave 1 +Slave 2 Chassis | Max. 6 channels with valves | 
| Aux.oven | H2 trap only for N2 measurement | 
| Carrier Gas | Helium | 
| Power | 115 VAC, 50-60 Hz | 
| 220 VAC, 50-60 Hz | |
| Gas Connectors (to match accessories fittings) | 1/8" VCR | 
| Serial port | RS-232 | 
| Remote Support | VNC | 
| Net Support | |
| Internal Dilution System | Not Required | 
| Required ( One per Chassis) | |
| 04402A1 | Argon Purity Analyser Base | 
| includes report software, 2 SS coils tubing, packaging | |
| Trace (>10 ppm), Pack 1 | O2 Impurity - Enter range & LDL | 
| 0-10ppm ≤O2 range ≤ 0-100ppm LDL is 1.5% full scale | |
| H2 Impurity - Enter range & LDL | |
| 0-10ppm ≤H2 range ≤ 0-100ppm LDL is 1.5% full scale | |
| N2 Impurity - Enter range & LDL | |
| 0-10ppm ≤N2 range ≤ 0-100ppm LDL is 1.5% full scale | |
| CH4 Impurity - Enter range & LDL | |
| 0-10ppm ≤CH4 range ≤ 0-100ppm LDL is 1.5% full scale | |
| CO Impurity - Enter range & LDL | |
| 0-10ppm ≤CO range ≤ 0-100ppm LDL is 1.5% full scale | |
| Trace (>10 ppm), Pack 2 | Same as Pack 1 + | 
| CO2 Impurity - Enter range & LDL | |
| 0-10ppm ≤CO2 range ≤ 0-100ppm LDL is 1.5% full scale | |
| Trace Exact (<10 ppm), Pack 1 | O2 Impurity - Enter range & LDL | 
| 0-1ppm ≤O2 range < 0-10ppm LDL is 1.5% full scale,Min.50ppb | |
| Min Range 0-1 ppm | H2 Impurity - Enter range & LDL | 
| 0-1ppm ≤H2 range < 0-10ppm LDL is 1.5% full scale,Min.50ppb | |
| N2 Impurity - Enter range & LDL | |
| 0-1ppm ≤N2 range < 0-10ppm LDL is 1.5% full scale,Min.50ppb | |
| CH4 Impurity - Enter range & LDL | |
| 0-1ppm ≤CH4 range < 0-10ppm LDL is 1.5% full scale,Min.50ppb | |
| Trace Exact (<10 ppm), Pack 2 | Same as Pack 1 + | 
| Min Range 0-1 ppm | CO Impurity - Enter range & LDL | 
| 0-1ppm ≤CO range < 0-10ppm LDL is 1.5% full scale,Min.50ppb | |
| Trace Exact (<10 ppm), Pack 3 | Same as Pack 2 + | 
| CO2 Impurity - Enter range & LDL | |
| Min Range 0-1 ppm | 0-1ppm ≤CO2 range < 0-10ppm LDL is 1.5% full scale,Min.50ppb | 
| Carrier Gas | Argon | 
| Power | 115 VAC, 50-60 Hz | 
| 220 VAC, 50-60 Hz | |
| Gas Connectors (to match accessories fittings) | 1/8" Swagelok | 
| 1/8" VCR (if 0-1 ppm range) | |
| Serial port | Not Required | 
| RS-232 | |
| Remote Support | VNC | 
| Net Support | |
| Dual Inlets (sample + span, with switching) | Not Required | 
| Required | |
| Calibration using background gas | Not required | 
| Required | |
| 04403A1 | Helium Purity Analyser Base | 
| includes report software, SS liners, packaging | |
| Trace (>10 ppm), Pack 1 | O2 Impurity - Enter range & LDL | 
| 0-10ppm ≤O2 range ≤ 0-100ppm LDL is 1.5% full scale | |
| H2 Impurity - Enter range & LDL | |
| 0-10ppm ≤H2 range ≤ 0-100ppm LDL is 1.5% full scale | |
| N2 Impurity - Enter range & LDL | |
| 0-10ppm ≤N2 range ≤ 0-100ppm LDL is 1.5% full scale | |
| CH4 Impurity - Enter range & LDL | |
| 0-10ppm ≤CH4 range ≤ 0-100ppm LDL is 1.5% full scale | |
| CO Impurity - Enter range & LDL | |
| 0-10ppm ≤CO range ≤ 0-100ppm LDL is 1.5% full scale | |
| Trace (>10 ppm), Pack 2 | Same as Pack 1 + | 
| Ar Impurity - Enter range & LDL | |
| 0-10ppm ≤Ar range ≤ 0-100ppm LDL is 1.5% full scale | |
| Ne Impurity - Enter range & LDL | |
| 0-10ppm ≤Ne range ≤ 0-100ppm LDL is 1.5% full scale | |
| Trace (>10 ppm), Pack 3 | Same as Pack 2 + | 
| CO2 Impurity - Enter range & LDL | |
| 0-10ppm ≤CO2 range ≤ 0-100ppm LDL is 1.5% full scale | |
| Trace Exact (<10 ppm), Pack 1 | O2 Impurity - Enter range & LDL | 
| 0-1ppm ≤O2 range < 0-10ppm LDL is 1.5% full scale,Min.50ppb | |
| Min Range 0-1 ppm | H2 Impurity - Enter range & LDL | 
| 0-1ppm ≤H2 range < 0-10ppm LDL is 1.5% full scale,Min.50ppb | |
| N2 Impurity - Enter range & LDL | |
| 0-1ppm ≤N2 range < 0-10ppm LDL is 1.5% full scale,Min.50ppb | |
| CH4 Impurity - Enter range & LDL | |
| 0-1ppm ≤CH4 range < 0-10ppm LDL is 1.5% full scale,Min.50ppb | |
| Trace Exact (<10 ppm), Pack 2 | Same as Pack 1 + | 
| CO Impurity - Enter range & LDL | |
| Min Range 0-1 ppm | 0-1ppm ≤CO range < 0-10ppm LDL is 1.5% full scale,Min.50ppb | 
| Trace Exact (<10 ppm), Pack 3 | Same as Pack 2 + | 
| Min Range 0-1 ppm | Ar Impurity - Enter range & LDL | 
| 0-1ppm ≤Ar range < 0-10ppm LDL is 1.5% full scale,Min.50ppb | |
| Ne Impurity - Enter range & LDL | |
| 0-1ppm ≤Ne range < 0-10ppm LDL is 1.5% full scale,Min.150ppb | |
| Trace Exact (<10 ppm), Pack 4 | Same as Pack 3 + | 
| CO2 Impurity - Enter range & LDL | |
| Min Range 0-1 ppm | 0-1ppm ≤CO2 range < 0-10ppm LDL is 1.5% full scale,Min.50ppb | 
| Carrier Gas | Helium | 
| Power | 115 VAC, 50-60 Hz | 
| 220 VAC, 50-60 Hz | |
| Gas Connectors (to match accessories fittings) | 1/8" Swagelok | 
| 1/8" VCR (if 0-1 ppm range) | |
| Serial port | Not Required | 
| RS-232 | |
| Remote Support | VNC | 
| Net Support | |
| Dual Inlets (sample + span, with switching) | Not Required | 
| Required | |
| Calibration using background gas | Not required | 
| Required | |
 
        
                     
         
         
         
         
         
                            

 
             
            